Preparation
Operators must wear acid-resistant gloves and safety goggles, and perform operations inside a fume hood. Identify the types of contaminants present (e.g., organic matter, inorganic salts, metal ions) beforehand and prepare the appropriate reagents and tools.
Pre-treatment and Initial Rinsing
Wipe the outer wall with a lint-free cloth dampened with ethanol to avoid contaminating the inner cavity. Purge loose dust from the tube interior using low-pressure nitrogen (0.2–0.5 MPa) to prevent particles from scratching the inner wall.
Circulate hot deionized water (above 70°C) through the tube for 10 minutes to soften loosely adhered deposits and reduce electrostatic attraction.
Post-treatment and Quality Verification
Use isopropyl alcohol (IPA) vapor for drying instead of conventional oven drying to prevent residual moisture from causing new contamination.
Measure light transmittance at a wavelength of 400 nm using a UV-Vis spectrophotometer; the value should be ≥92%. A contact angle of θ < 10° indicates good hydrophilicity, meeting cleanliness requirements.
Seal the tube openings and tail sections with specialized aluminum foil and store them properly in a designated area within the cleanroom to prevent re-contamination.